Chemical pathology case conference - laboratory tests for diabetes mellitus
Yuet-ping Yuen 袁月冰, Chloe M Mak 麥苗, Angel O K Chan 陳安琪, Michael H M Chan 陳浩明, Rossa
W K Chiu 趙慧君, Ching-wan Lam 林青雲, Tony W L Mak 麥永禮, Wing-tat Poon 潘永達, Anthony C
C Shek 石志忠, Morris H L Tai 戴學良, Sidney Tam 譚志輝, Albert Y W Chan 陳恩和
HK Pract 2007;29:419-426
Summary
Diabetes mellitus is a chronic disease with increasing prevalence in most places
in the world. A number of laboratory tests, such as glucose, haemoglobin A1C, urine
albumin, creatinine, lipid profile play pivotal roles in diabetes care. Other adjunctive
laboratory tests include urine and blood ketone, fructosamine, and C-peptide. Judicious
use of these tests is important to ensure all patients are diagnosed and managed
according to evidence-based clinical guidelines. Some knowledge about the pre-analytical
and analytical aspects and potential pitfalls of these laboratory tests also assist
clinicians to interpret test results properly.
摘要
糖尿病是一種慢性疾病,在世界大多數地區,它的病發率都在上升。血糖、血紅蛋白A1C、尿蛋白、肌酐、 血脂等化驗檢查在治療糖尿病時扮演著重要的角色。其他輔助化驗項目還有尿酮體和血酮體、果糖胺和C。 明智地選擇適當的檢查可確保病人能按照有實證基礎的臨床指引得到診治。能夠認識化驗前和化驗時的樣本處理過程,
以及它們的潛在問題,也有助醫生正確地詮釋化驗結果。
Introduction
Diabetes mellitus is a chronic disease with increasing prevalence in most places
in the world. In Hong Kong, it is estimated that 1 in 10 people have diabetes.1
There are a number of national and international clinical guidelines for diabetes
mellitus. In particular, the American Diabetes Association (ADA) and the World Health
Organization (WHO) guidelines are widely adopted by health care providers globally,
including Hong Kong. Table 1 lists the diagnostic criteria of diabetes and
pre-diabetes from ADA and WHO. The Diabetes Division, Hong Kong Society for Endocrinology,
Metabolism and Reproduction also published a statement on the clinical diagnosis
and management for type 2 diabetes mellitus in year 2000.2 To implement
any one of these guidelines in clinical practice, the use of multiple laboratory
tests is necessary. This article focuses on the pre-analytical and analytical aspects
of laboratory tests that are used in diagnosis and management of diabetes. Readers
can refer to the published clinical guidelines for other aspects of diabetes care
like classification, management and treatment targets.
Laboratory tests for diagnosis
The diagnosis of diabetes depends on the measurement of fasting plasma glucose (FPG)
or the 2-hour plasma glucose after an oral glucose tolerance test (OGTT).3,4
A random plasma glucose level can also be used for diagnosis if patients have symptoms
like polydipsia, polyuria and weight loss. Other tests like glycated haemoglobin
(GHb) is currently not recommended for this purpose.3,4 Before taking
a blood sample for FPG, patients have to be fasted for at least eight hours.4
Fasting plasma glucose has a diurnal variation, with mean level higher in the morning
than in the afternoon.5 Therefore, performing the test in the afternoon
may miss some patients with diabetes.6 Glucose concentrations in blood
samples decline continuously after blood collection because of glycolysis. Antiglycolytics
like sodium fluoride (NaF) is commonly used to minimize this ex vivo change. If
cells can be separated within an hour after blood collection, antiglycolytics are
not required and heparin plasma can be used.
Glucose can be measured in plasma and whole blood collected from different sites
(venous, capillary and arterial). Glucose concentration in plasma is approximately
11% higher than that of whole blood because of higher water concentration in plasma.
However, this difference in glucose concentration would vary with changes in the
haematocrit level.7 Because of this correlation problem between plasma
and whole blood glucose, both ADA and WHO recommend the use of venous plasma glucose
for diagnosis of diabetes. All glucose measurements used for diagnosis of diabetes
have to be performed in a clinical laboratory using a laboratory analytical method.
Glucometers, which has limited accuracy and precision at both high and low glucose
concentrations, is not suitable. Because of the relatively large intra-individual
biological variations, repeat testing is required to confirm a diagnosis.
Laboratory tests for determining the aetiology of diabetes
Immune-mediated type 1 diabetes is caused by autoimmune destruction of pancreatic
-cells, leading eventually to absolute
insulin deficiency. One or more autoimmune markers can be detected in patients with
immune-mediated diabetes. These markers include cytoplasmic islet cell antibodies
(ICAs, detected by indirect immunofluorescence on frozen section of pancreatic cells),
insulin autoantibodies (IAA), autoantibodies to glutamic acid decarboxylase (GAD)
and autoantibodies to two tyrosine phosphatases (IA-2A and 1A-2bA).8
A subset of patients with type 1 diabetes does not have evidence of autoimmunity.
They are known as idiopathic diabetes. Type 2 diabetes is a disease with combined
insulin resistance and relative insulin deficiency. In Hong Kong, the majority of
diabetic patients have type 2 diabetes.9
Most patients are classified as type 1 or type 2 diabetes on the basis of age of
disease onset, mode of presentation, insulin dependence, family history and body
mass index. Specific diagnostic tests like autoimmune markers are rarely necessary.6
C-peptide is used as a marker of
-cell reserve. Both fasting and
post-glucagon stimulation levels have been used in differentiating type 1 from type
2 diabetes.10,11 However, these tests should be reserved for patients
who required a definitive classification and should be performed under specialist
care.
Hormones including cortisol, growth hormones, glucagon and adrenaline antagonize
the actions of insulin. Therefore, some endocrine diseases like Cushing syndrome,
acromegaly, glucagonoma and phaeochromocytoma can cause secondary hyperglycaemia.
Both hyperthyroidism and hypothyroidism affect diabetes management. Hyperthyroidism
can worsen glycaemic control and even precipitate diabetic ketoacidosis (DKA). On
the other hand, patients with hypothyroidism are at increased risk of hypoglycaemia
and more severe dyslipidaemia. Signs and symptoms of these endocrine diseases should
be looked for in newly diagnosed diabetic patients. If there is any clinical suspicion,
specific laboratory investigations should be performed.
Mitochondrial DNA mutations like 3243A>G can cause maternally inherited diabetes.
High incidence of deafness is found in this group of patients.12 Genetic
tests for hotspot mutations in mitochondrial DNA (mtDNA) are now available as routine
tests in some local clinical laboratories. Patient with clinical features of mitochondrial
disorders (e.g. maternal inheritance, deafness, short stature, myopathy and any
neurological complaints) should be referred for genetic testing.13 Another
genetic form of diabetes is maturity onset diabetes of the young (MODY). At least
eight different types of MODY have been characterized at the molecular level and
all of them are inherited in an autosomal dominant manner. Genetic tests for MODY
is still largely performed in research settings and no routine service is available
in local laboratories.
Laboratory tests for complication detection
Proteinuria
Diabetic nephropathy is one of the most common causes of end-stage renal disease
and its first manifestation is increased urine albumin excretion.14 Albuminuria
of diabetic nephropathy is divided into two stages - microalbuminuria and macroalbuminuria
(Table 2). The use of 24-hour or timed urine collection is the traditional
way for diagnosis of microalbuminuria. However, spot urine sample is now recommended
for urine albumin measurement and the result is expressed as albumin-to-creatinine
ratio (ACR) (units: mg/mmol creatinine or mg/g creatinine) or simply as albumin
concentration (mg/L).14-16 The use of spot urine is convenient to patients
and it also eliminates collection error that is common in 24-hour urine collection.
Different cut-offs of ACR are used to define micro- and macro-albuminuria in different
clinical guidelines.14,16 The cut-off values that are commonly used by
local clinicians are listed in Table 2. The use of creatinine in ACR is to
correct for variations caused by hydration. However, urine creatinine concentration
is determined by factors like age, sex and muscle mass. Therefore, some advocate
the use of sex-specific ACR cut-offs (> 2.5 mg/mmol creatinine for males and > 3.5
mg/mmol creatinine for females) as female subjects have lower urine creatinine concentration
than male subjects.16,17
First morning urine is the preferred sample for ACR measurement but random urine
sample is also acceptable.14,16 However, significant discrepancy between
ACR from first morning urine and random spot urine has been reported.18
Therefore, consistent time of urine collection may make serial ACR results of a
patient more comparable. Interpretation of results may not be straight-forward when
in the presence of the following conditions that may increase urine albumin excretion:
urinary tract infection, haematuria, acute febrile illness, vigorous exercise, short-term
pronounced hyperglycaemia, uncontrolled hypertension and congestive heart failure.14
Most clinical laboratories measure urine albumin by commercial assays using either
immunoturbidimetry or immunonephelometry. The detection limits and precision of
these methods allow the accurate measurement of urine albumin at low concentrations.6
Different dipsticks are available for measurement of urine proteins. Standard urine
dipsticks (e.g. Albustix) detects total protein in urine using pH indicator dye
which changes colour upon binding to negatively charged proteins. These dipsticks
are not sensitive enough for detection of microalbuminuria. More sensitive dipsticks
for qualitative or semiquantitative measurements of microalbuminuria are also available.
These microalbumin dipsticks have acceptable sensitivity and specificity for detection
of albumin concentration at 20 mg/L, a cut-off used in some national guidelines
for microalbuminuria.16 However, their performance is inferior to ACR
measured by laboratory methods.19,20 So when laboratory tests for microalbumin
are not readily available, microalbumin dipsticks can be used. However, all positive
results have to be confirmed by a laboratory test.
All type 2 diabetic patients should be screened for diabetic nephropathy at initial
assessment. For type 1 diabetic patients, screening may start five years after disease
onset.14 A spot urine sample should be first checked for the presence
of protein using standard urine dipstick. If it is negative, a test for either ACR
(spot urine) or urine albumin excretion rate (UAER, 24-hour or timed urine) should
follow. Because of the large intra-individual biological variation, repeated measurements
of ACR or UAER are required to confirm the diagnosis.14 On the other
hand, if a spot urine is tested positive for protein by standard urine dipstick,
overt nephropathy is present. Then urine total protein should be measured.
Renal function and glomerular filtration rate (GFR)
In diabetic patients, assessment of glomerular function is as important as albuminuria
screening. Significant decline in GFR has been noted in patients with type 1 and
type 2 diabetes and normal urine albumin excretion.14 Therefore, screening
for increased urine albumin alone would miss a considerable number of diabetic patients
with chronic kidney disease. However, plasma creatinine level is determined by many
factors other than GFR and it is not sensitive for detection of mild decrease in
GFR. Other methods of GFR assessment for routine clinical use include calculation
of creatinine clearance (CrCl) using 24-hour urine samples and the use of estimation
equations based on plasma creatinine and other parameters. More detailed discussions
about the merits and problems of different GFR assessment methods can be found in
a previous Chemical Pathology Case Conference.21
Lipid profile
Elevated triglyceride with low HDL-cholesterol is commonly seen in diabetic patients.
As both diabetes and hyperlipidaemia are risk factors for cardiovascular diseases,
all newly diagnosed diabetic patients should have a fasting lipid profile checked.
Severe hypertriglyceridaemia (>30 mmol/L) can be seen in patients with recent onset
diabetes. Blood samples from these patients appear turbid to naked eyes.
Laboratory tests for monitoring of glycaemic control
Glucose
Self-monitoring of blood glucose (SMBG) using glucometers plays an important role
in routine monitoring of diabetes. As mentioned above, glucose concentrations in
plasma and whole blood are different. Some glucometers measure glucose concentrations
in whole blood sample and are programmed to report plasma glucose concentrations
but some glucometers still report whole blood glucose values.3,22 Therefore,
it is important to check the package inserts of glucometers. Glucose concentrations
in capillary and venous whole blood are similar in fasting samples. However, a difference
of 20 - 25% (capillary higher than venous) can be seen in postprandial samples.23
Glucometer readings are not only used to monitor daily glycaemic control, they also
guide the dosages of insulin and help detect hypoglycemia. Proper quality control
of glucometer is thus crucial. The newer models of glucometers are more user-friendly
and require less skill to perform a test. Nonetheless, patients should be instructed
the correct use of glucometers regularly. Calibration and quality control should
be performed according to the manufacturer"s instructions. Periodic parallel test
with laboratory glucose analysis will help monitor the performance of the glucometers.
Moreover, test strips should be stored properly and all expired test strips should
not be used. Faulty strips may give erroneous results that lead to serious consequences.
Glucometers are usually calibrated to cater for a normal haematocrit of 40 - 50%.
They give lower glucose results when the haematocrit is above this range and higher
glucose results when the haematocrit is below this range.24 However,
this would not be a significant problem in most diabetic patients as their haematocrit
levels would not deviate much from the normal range. Likewise, other factors that
have been reported to influence glucometers readings (e.g. partial pressure of oxygen
in blood, the presence of sugars like maltose in blood, etc.) have little impact
on ordinary diabetic patients.25,26
Urine glucose testing, once commonly used in home glucose monitoring, has been replaced
by SMBG.23 Urine glucose testing is easy to perform and convenient to
patients, but its clinical usefulness is limited as it provides no information about
blood glucose levels below the variable renal threshold (~ 10 mmol/L). It reflects
only the average glucose values between voiding and a normal result cannot distinguish
euglycaemia from hypoglycaemia. Together with other deficiencies like poor performance
at low glucose levels, urine glucose testing should only be relied upon when there
is no better alternative.23
Glycated haemoglobin (GHb)
Glycated haemoglobins (GHb) are formed by non-enzymatic reactions between glucose
and the amino acid residues of haemoglobin (Hb) molecules, and its level is dependent
on the blood glucose concentration. Therefore, GHb level is used as an index of
average glycaemia in the previous 2 - 3 months (i.e. the average lifespan of red
blood cells).23 Haemlglobin A1C (Hb A1C) is one of the many species of
GHb exist in the circulation. It is a specific GHb with a glucose attached to the
NH2-terminal valine of one or both b-chains. GHb is the standard for monitoring
glycaemic control in diabetic patients because its concentration in blood is linked
to the risks of development of diabetic complications as shown by the Diabetic Control
and Complications Trials (DCCT) and the United Kingdom Prospective Diabetes Studies
(UKPDS).27,28 To adopt the results of these prospective studies, HbA1C
have to be measured by assays that are certified to produce DCCT-aligned values.
More information about the certified HbA1C methods can be found in the National
Glycohemoglobin Standardization Program (NGSP) website (www.ngsp.org).
A consensus statement on the worldwide standardization of HbA1C measurement
was published recently, which suggested standardizing of all HbA1C assays
to the International Federation of Clinical Chemistry (IFCC) reference system and
reporting HbA1C results in mmol/mol in addition to the traditional unit
(%).29 It is expected that most of the HbA1C assay manufacturers will
follow this new consensus guideline and a change in the HbA1C assays will occur
in the near future.
All diseases and conditions that affect red blood cell (RBC) survival or alter the
average age of circulating RBC affect HbA1C results irrespective of the
analytical methods used. For example, haemolytic anaemias decrease HbA1C
as RBC survival is shortened. Apart from this, HbA1C measurements are
also affected by Hb variants (e.g. HbS, HbC and HbF) and Hb derivatives (e.g. carbamylated
Hb which is elevated in patients with uraemia) in ways that are independent on their
effects on RBC survival.30 Moreover these effects are method dependent.
A summary of reported interferences of different Hb variants on different HbA1C
assays can be found in the NGSP web site. Hemoglobin F is the fetal hemoglobin with
two a and two g chains. The level of HbF usually falls to the adult level (< 1%)
at two years of age.31 The effects of HbF on HbA1C assays are variable
and most of the reported HbF interference occurs when HbF is higher than 10%.32
The most common haemoglobinopathies among Chinese are thalassemias. It has been
demonstrated that most commercial methods could measure HbA1C correctly in beta-thalassemia
trait samples.33-35 Iron status also affect the interpretation of HbA1C
results. It has been shown that in both diabetic and non-diabetic subjects with
co-existing iron deficiency, HbA1C levels dropped significantly after iron replacement.36
Fructosamine
Fructosamine, a collective term for serum glycated proteins, is less commonly used
than HbA1C in routine monitoring of glycaemic control. Serum proteins
(mainly albumin) have a circulating half-life of around 20 days and therefore fructosamine
can be used as an index of glycaemia over a period of 2 - 3 weeks. Assays for fructosamine
are not widely available in local clinical laboratories, probably related to the
small clinical demand. Fructosamine may be used in patients whose HbA1C levels do
not reliably reflect glycaemic control (e.g. patients with haemolytic anaemias).
Fructosamine levels are affected by any diseases and conditions that affect the
half-life of plasma proteins, such as nephrotic syndrome. The values are also unreliable
when there are gross changes or rapid changes in plasma protein concentrations.6
Urine and blood ketones
As recommended by ADA, urine ketone testing can be used in routine diabetic monitoring
especially for patients with type 1 diabetes and pregnant diabetic patients.23
All urine ketone testing devices detect acetoacetate, and some also detect acetone.
However, these devices cannot measure
-hydroxybutyrate, the major ketone
body formed during DKA. So urine ketone testing may provide misleading clinical
information by underestimating total ketone body concentration. Moreover, any free-sulfhydryl
compounds (e.g. captopril) can cause a false positive reaction in these devices.37
Devices for urine ketone testing can also be used for testing ketone in blood. POCT
devices for direct determination of b-hydroxybutyrate in whole blood are also available.38
These devices may be more sensitive than urine ketone testing in detecting incipient
DKA but further studies are required to confirm its clinical usefulness.
Conclusion
Quality care of diabetic patients depend heavily on use of laboratory tests, either
performed in a clinical laboratory, at the site of patient care or by patients themselves
as home monitoring. FPG and OGTT are the only tests for diagnosis of diabetes. All
glucose measurements for diagnosis should be measured by a laboratory analytical
method and not by glucometers. Specific diagnostic tests like autoimmune markers
and C-peptide for differentiating type 1 and type 2 diabetes are not necessary in
the majority of patients. With increasing knowledge of the underlying genetic causes
of diabetes and more readily assessable testing facilities, selected patients should
be referred for genetic testing. Certain endocrinopathies can cause secondary hyperglycaemia
or worsen glycaemic control. Appropriate endocrine tests should be performed if
there is a clinical suspicion. Regular screening for microalbuminuria using either
ACR or UAER is indicated for all diabetic patients. Dipstick for microalbuminuria
can be used when laboratory tests for microalbumin is not readily available but
any positive results have to be confirmed by a laboratory test. Glomerular functions
of diabetic patients may deteriorate without increase in urine albumin excretion.
Therefore, regular assessment of GFR is necessary. All diabetic patients should
have lipid profile checked regularly. Diabetes per se can lead to dyslipidaemia
and both diabetes and dyslipidaemia contribute to increased risk to cardiovascular
diseases. Glucometers are commonly used at the site of patient care and as home
monitoring. As the tests are performed outside laboratories, it is the responsibility
of the operators to perform regular calibration, quality control and maintenance
according to the manufacturer"s recommendations. A good quality control system should
be established to ensure accurate results from glucometers. Haemoglobin A1C plays
an indispensible role in monitoring glycaemic control. Global standardization would
improve the comparability of test results from different laboratories. However,
standardization alone cannot solve some of the inherent problems of HbA1C assays
like interference by Hb variants and Hb derivatives.
Key messages
- Diabetes can only be confirmed by measuring either fasting plasma glucose or 2-hour
plasma glucose after a 75-g oral glucose tolerance test using a laboratory glucose
assay. Glucometer readings cannot be used as a substitute.
- Different sample types (plasma vs. whole blood) and samples collected from different
sites (venous vs. capillary) can give different glucose results.
- Many clinical decisions are made on the basis of glucometer results. Any false positive
or false negative results could lead to disastrous consequences. Therefore, a good
quality control system with regular operator training and parallel testing with
laboratory glucose assay are essential to ensure accurate results.
- Spot urine sample with result expressed as albumin-to-creatinine ratio (ACR) can
be used to replace 24-hour or timed urine sample for microalbuminuria screening.
- Most existing commercial HbA1C assays are recalibrated to DCCT-aligned values. Global
standardization of HbA1C is underway. A new reference system and reporting unit
will be adopted.
- Haemoglobin variants and derivatives have differential effects on different HbA1C
assays. Global standardization would not overcome these method-specific problems.
Interferences should be suspected when HbA1C results do not match other laboratory
testing results.
Yuet-ping Yuen, MBChB(CUHK), FRCPA, FHKAM(Pathology)
Resident Specialist,
Albert Y W Chan, MBChB(Glasg), MD(CUHK), FHKCP, FHKCPath
Consultant Chemical Pathologist,
Department of Pathology, Princess Margaret Hospital.
Chloe M Mak, MBBS, FRCPA, FHKCPath, FHKAM(Pathology)
Resident Specialist,
Sidney Tam, FRCP(Edin), FRCPA, FHKAM(Medicine), FHKAM(Pathology)
Head and Consultant,
Division of Clinical Biochemistry, Department of Pathology, Queen Mary Hospital.
Wing-tat Poon, MBChB(CUHK), MRCP, FHKCPath, FHKAM(Pathology)
Resident Specialist,
Tony WL Mak, MBChB(CUHK), MBA, FRCPath, FRCPA, FHKAM(Pathology)
Consultant,
Hospital Authority Toxicology Reference Laboratory.
Angel OK Chan, BMedSc(Hons,CUHK), MBChB(CUHK), FRCPA, FHKAM(Pathology)
Resident Specialist,
Anthony CC Shek, MBBS(HK), FRCPath, FRCPA, FHKAM(Pathology)
Consultant,
Department of Pathology, Queen Elizabeth Hospital.
Michael HM Chan, MBChB(CUHK), FRCPA, FHKCPath, FHKAM(Pathology)
Associate Consultant,
Rossa WK Chiu, MBChB(Queensland), PhD(CUHK), FRCPA, FHKAM(Pathology)
Professor,
Ching-wan Lam, MBChB(CUHK), PhD(CUHK), FRCPA, FHKAM(Pathology)
Associate Professor,
Morris HL Tai, MBChB(CUHK), FRCPA, FHKAM(Pathology)
Medical Officer,
Department of Chemical Pathology, Prince of Wales Hospital, Chinese University of
Hong Kong.
Correspondence to : Dr Yuet-ping Yuen, Department of Pathology, Princess
Margaret Hospital, Kowloon, Hong Kong.
References
- Janus ED, Watt NM, Lam KS, et al. The prevalence of diabetes, association with cardiovascular
risk factors and implications of diagnostic criteria (ADA 1997 and WHO 1998) in
a 1996 community-based population study in Hong Kong Chinese. Hong Kong Cardiovascular
Risk Factor Steering Committee. American Diabetes Association. Diabet Med 2000;17:741-745.
- A statement for health care professionals on type 2 diabetes mellitus in Hong Kong.
Diabetes division, Hong Kong Society for Endocrinology, Metabolism, and Reproduction.
Hong Kong Med J 2000;6:105-107.
- World Health Organization. Definition and diagnosis of diabetes mellitus and intermediate
hyperglycemia. Report of a WHO/IDF consultation. 2006. http://www.who.int/diabetes/publications/Definition%20and%20diagnosis%20of%20diabetes_new.pdf
. (last assessed 5 October 2007).
- American Diabetes Association. Standards of medical care in diabetes -2007. Diabetes
Care 2007;30 Suppl 1:S4-S41.
- Troisi RJ, Cowie CC, Harris MI. Diurnal variation in fasting plasma glucose: implications
for diagnosis of diabetes in patients examined in the afternoon. JAMA 2000;284:3157-3159.
- Sacks DB, Bruns DE, Goldstein DE, et al. Guidelines and recommendations for laboratory
analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002;48:436-472.
- D"Orazio P, Burnett RW, Fogh-Andersen N, et al. Approved IFCC recommendation on
reporting results for blood glucose (abbreviated). Clin Chem 2005;51:1573-1576.
- American Diabetes Association. Diagnosis and classification of diabetes mellitus.
Diabetes Care 2007;30 Suppl 1:S42-47.
- Chan JC. Heterogeneity of diabetes mellitus in the Hong Kong Chinese population.
The Chinese University of Hong Kong-Prince of Wales Hospital Diabetes Research and
Care Group. Hong Kong Med J 2000;6:77-84.
- Service FJ, Rizza RA, Zimmerman BR, et al, 3rd. The classification of diabetes by
clinical and C-peptide criteria. A prospective population-based study. Diabetes
Care 1997;20:198-201.
- Tung YC, Lee JS, Tsai WY, et al. Evaluation of beta-cell function in diabetic Taiwanese
children using a 6-min glucagon test. Eur J Pediatr 2007 (in press).
- Kadowaki T, Kadowaki H, Mori Y, et al. A subtype of diabetes mellitus associated
with a mutation of mitochondrial DNA. N Engl J Med 1994;330:962-968.
- Ng MC, Lee SC, Ko GT, et al. Familial early-onset type 2 diabetes in Chinese patients:
obesity and genetics have more significant roles than autoimmunity. Diabetes Care
2001;24:663-671.
- American Diabetes Association. Nephropathy in diabetes. Diabetes Care 2004;27 Suppl
1:S79-83.
- K/DOQI clinical practice guidelines for chronic kidney disease: evaluation, classification,
and stratification. Am J Kidney Dis 2002;39:S1-266.
- National Institute for Health and Clinical Excellence. Management of type 2 diabetes
- renal disease, prevention and early management. 2002.
- http://guidance.nice.org.uk/page.aspx?o=27924 . (last assessed 5 October 2007).
Houlihan CA, Tsalamandris C, Akdeniz A, et al. Albumin to creatinine ratio: a screening
test with limitations. Am J Kidney Dis 2002;39:1183-1189.
- Babazono T, Takahashi C, Iwamoto Y. Definition of microalbuminuria in first-morning
and random spot urine in diabetic patients. Diabetes Care 2004;27:1838-1839.
- Incerti J, Zelmanovitz T, Camargo JL, et al. Evaluation of tests for microalbuminuria
screening in patients with diabetes. Nephrol Dial Transplant 2005;20:2402-2407.
- Parikh CR, Fischer MJ, Estacio R, et al. Rapid microalbuminuria screening in type
2 diabetes mellitus: simplified approach with Micral test strips and specific gravity.
Nephrol Dial Transplant 2004;19:1881-1885.
- Yuen YP, Tam S, Chan AKC, et al. Chemical pathology case conference - renal function
tests. HK Pract 2006;28:115-122.
- Ghys T, Goedhuys W, Spincemaille K, et al. Plasma-equivalent glucose at the point-of-care:
evaluation of Roche Accu-Chek Inform(R) and Abbott Precision PCx(R) glucose meters.
Clin Chim Acta 2007;386:63-68.
- Goldstein DE, Little RR, Lorenz RA, et al. Tests of glycemia in diabetes. Diabetes
Care 2004;27:1761-1773.
- Tang Z, Lee JH, Louie RF, et al. Effects of different hematocrit levels n glucose
measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med
2000;124:1135-1140.
- Schleis TG. Interference of maltose, icodextrin, galactose, or xylose with some
blood glucose monitoring systems. Pharmacotherapy 2007;27:1313-1321.
- Tang Z, Louie RF, Lee JH, et al. Oxygen effects on glucose meter measurements with
glucose dehydrogenase- and oxidase-based test strips for point-of-care testing.
Crit Care Med 2001;29:1062-1070.
- The effect of intensive treatment of diabetes on the development and progression
of long-term complications in insulin-dependent diabetes mellitus. The Diabetes
Control and Complications Trial Research Group. N Engl J Med 1993;329:977-986.
- Intensive blood-glucose control with sulphonylureas or insulin compared with conventional
treatment and risk of complications in patients with type 2 diabetes (UKPDS 33).
UK Prospective Diabetes Study (UKPDS) Group. Lancet 1998;352:837-853.
- Consensus statement on the worldwide standardization of the hemoglobin A1C measurement:
the American Diabetes Association, European Association for the Study of Diabetes,
International Federation of Clinical Chemistry and Laboratory Medicine, and the
International Diabetes Federation. Diabetes Care 2007;30:2399-2400.
- Bry L, Chen PC, Sacks DB. Effects of hemoglobin variants and chemically modified
derivatives on assays for glycohemoglobin. Clin Chem 2001;47:153-163.
- Rochette J, Craig JE, Thein SL. Fetal hemoglobin levels in adults. Blood Rev 1994;8:213-224.
- Rohlfing C, Connolly S, England J, et al. Effect of Elevated Fetal Hemoglobin on
HbA1c Measurements: Four Common Assay Methods compared to the IFCC Reference Method.
AACC Annual Conference, Chicago, IL. 2006.
- Piras G, Carluccio A, Coe A, et al. Monitoring of HbA(1C) in patients with thalassemia
and sickle cell disease. Klinisches Labor 1993;39:1033-1037.
- Polage C, Little RR, Rohlfing CL, et al. Effects of beta thalassemia minor on results
of six glycated hemoglobin methods. Clin Chim Acta 2004;350:123-128.
- Weykamp CW, Penders TJ, Muskiet FA, et al. Influence of hemoglobin variants and
derivatives on glycohemoglobin determinations, as investigated by 102 laboratories
using 16 methods. Clin Chem 1993;39:1717-1723.
- Tarim O, Kucukerdogan A, Gunay U, et al. Effects of iron deficiency anemia on hemoglobin
A1c in type 1 diabetes mellitus. Pediatr Int 1999;41:357-362.
- Csako G.. Causes, consequences, and recognition of false-positive reactions for
ketones. Clin Chem 1990;36:1388-1389.
- Chiu RW, Ho CS, Tong SF, Ng KF, et al. Evaluation of a new handheld biosensor for
point-of-care testing of whole blood beta-hydroxybutyrate concentration. Hong Kong
Med J 2002;8:172-176.
- Wat NM, Lam KS. Clinical management manual on type 2 DM in the Western Cluster.
Hong Kong. Diabetes Centre, Queen Mary Hospital, Department of Medicine, University
of Hong Kong 2001.
|